 Compact Medium Power Amplifier for EMI/EMC Test and Measurement Applications
4 . 0 - 1 0 .0 GHz
The VZS/X-2776L1 Series 100 W TWT Amplifier
4.0 - 10.0 GHz
1 0 0 W Tr a v e l i n g Wa v e Tu b e A m p l i f i e r
Compact
Global Applications
Three rack units tall (5.25 in, 133 mm)
Versatile
230 VAC operation. Meets International Safety Standard EN61010 and Electromagnetic Compatibility 2004/108/EC.
Worldwide Support
Ultra-wideband, automatic fault recycle, userfriendly microprocessor-controlled logic with integrated computer interface, VSWR soft-fail protection, digital metering, quiet operation for laboratory environment. An integral solid state pre-amplifier and IEEE interface are included as standard features.
Efficient
Modular design and built-in fault diagnostic capability, backed by CPI's worldwide 24-hour customer support network that includes sixteen regional factory service centers.
Utilizes dual-depressed collector helix traveling wave tube for maximum 1.5 kVA operation.
45 River Drive Georgetown, Ontario, Canada L7G 2J4 tel: +1 (905) 877-0161 fax: +1 (905) 877-5327 e-mail: marketing@satcom.cpii.com www.cpii.com/satcom
4 .0 - 10 .0 G H z
SPECIFICATIONS, VZS/X-2776L1
Electrical
Frequency 4.0 to 10.0 GHz 100 W min. 70 W min. 44 dB min. at rated power output; 40 dB typ. at small signal 0 to 20 dB continuous 0.25 dB/24 hr max. (at constant drive and temp.) 12.0 dB pk-pk max., typ. 2.5:1 typ. 1.5:1 max. (with optional input isolator) 2.5:1 typ. 1.5:1 max. for full spec compliance; 2.0:1 max. continuous operation; any value without damage -50 dBc below 10 kHz; -20 [1.3 + log F (kHz)] dBc, 10 kHz to 500 kHz; -80 dBc above 500 kHz Meets IESS-308/309 with 3 dB margin 15 dB max. -3 dBc typical at lower band edge, decreasing to -15 dBc typ. at upper band edge 220-240 VAC 10%, single phase, 47-63 Hz 1.4 kVA typ. 1.5 kVA max. 200% max. Output Power TWT Flange Gain (with no options) RF Level Adjust Gain Stability Gain Variation Input VSWR Output VSWR Load VSWR
Environmental (Operating)
Ambient Temperature Relative Humidity Altitude Shock and Vibration Acoustic Noise -10 to +40C operating 95% non-condensing Up to 10,000 ft (3000 m) with standard adiabatic derating of 2/1000 ft. Designed to meet conditions normally encountered in the laboratory 65 dBA @ 3 ft. from amplifier
OPTIONS: * Input Isolator (-1 dB gain) * Remote Control Panel * 115 VAC External Step-Up Transformer
Mechanical
Cooling (TWT) RF Input Connection RF Output Connection RF Output Monitor Dimensions (W x H x D) Weight Safety Forced air with integral blower. Rear air intake and exhaust. Type N female Type N female Type-N female, -57 dB nom. 19 x 5.25 x 24 in. (483 x 133 x 610 mm) 70 lbs (32 kg) EN61010
Residual AM
Phase Noise Noise Figure Harmonic Content
Primary Power Power Consumption Inrush Current
Certificate Number: 30515
Note: Specifications may change without notice as a result of additional data or product refinement. Please contact CPI before using this information for system design. MKT 234, ISSUE 1 03/08 PDF


▲Up To Search▲   

 
Price & Availability of VZS-2776L1

All Rights Reserved © IC-ON-LINE 2003 - 2022  

[Add Bookmark] [Contact Us] [Link exchange] [Privacy policy]
Mirror Sites :  [www.datasheet.hk]   [www.maxim4u.com]  [www.ic-on-line.cn] [www.ic-on-line.com] [www.ic-on-line.net] [www.alldatasheet.com.cn] [www.gdcy.com]  [www.gdcy.net]


 . . . . .
  We use cookies to deliver the best possible web experience and assist with our advertising efforts. By continuing to use this site, you consent to the use of cookies. For more information on cookies, please take a look at our Privacy Policy. X